Amain AMOX-5A Cheap Health Care Use Outdoor Use Oxygen-Concentrator With Atomize Function For Sale
Characteristics
Error Alarm Purity Detection
Four alarm functions, let you use more secure and assured.
LOW OXYGEN PURITY ALARM
The L.O2. light will be red when the oxygen purity is less than 82%.
HIGH TEMPERATURE ALARM
When temp in machine exceed 50℃, H.T. light will be red while with intermittent alarm sound.
HIGH&LOW PRESSURE ALARM
When the machine running pressure is too high, the H.P. light will be red, When the running pressure is too low, the L.P. light will be yellow; both will be with continuously alarm sound.
POWER FAILURE ALARM
P.F. light is red when power is off.
Specification
Model
|
JAY-5BW
|
Flow rate
|
0-5L/min
|
Oxygen purity
|
93±3%
|
Outlet pressure
|
0.04-0.07Mpa
|
Noise level
|
≤43db
|
Power Supply
|
AC230V, 50Hz; AC220V/110V (±10%), 50/60Hz (±1Hz)
|
Power consumption
|
≤540W
|
LCD display
|
Switch times, operating pressure, present working time, accumulated working time, preset time from 10mins to 40hours
|
Alarm
|
Power failure alarm
|
Size
|
360x300x600mm
|
Net weight
|
23kg
|
Opitional configuration |
1.Nebulizer (Atomization): >10L/min
2.SPO2(Pulse oximeter) 3.Low purity alarm:when oxygen purity is above 82% ,it will give blue light,when the purity is below 82%(82% not included),it will give red light 4.High temperature alarm(Inside system temp above 50℃) |
Product Configuration
Standard configuration:
Air Intake Filtering Net, Air Intake Filtering Cloth, Power Cord, User Manual, Main Unit, Nasal Cannula