Product Description
AMAIN Automatic Biochemistry Analyzer AMBC3000 For Laboratory In Vitro Diagnostic Analyzer
Image Gallery
Specification
Light source
|
Halogen-tungsten lamp, 12V 20W
|
||
Temperature control
|
37°C;
|
||
Temperature accuracy
|
±0.1°C;
|
||
Reaction disc
|
81 cuvettes, optical length 5mm;
|
||
Cuvette
|
Semi-permanent plastic cuvette, automatic cleaning, recycling;
|
||
Wavelength
|
Standard 8 wavelengths: 340nm-670nm; customized wavelength available;
|
||
Stray light
|
Absorbance ≥2.5
|
||
Absorbance linearity
|
The maximum absorbance of relative deviation within ±2% is no less than 4.5
|
||
Absorbance accuracy
|
The error is no more than ±0.015 when the absorbance is 0.5, and the error is no more than ±0.025 when the absorbance is 1.0
|
||
Absorbance stability
|
The change of absorbance is no more than 0.005;
|
||
Absorbance repeatability
|
CV≤1.0%;
|
||
Temperature accuracy and fluctuation
|
Temperature value within 37°C±0.2°C, the fluctuation is no more than ±0.1°C;
|
||
Sample carry-over rate
|
≤0.5%
|
||
Sampling accuracy and CV
|
Accuracy≤±5%, CV≤2%
|
||
Within-run precision (CV%)
|
ALT≤5%; TP≤2.5%; UREA≤2.5%
|
||
Data storage
|
Automatic storage, automatic backup, permanent storageof various types of data
|
||
Power supply
|
AC100V-240V, 50/60Hz
|
Product Application
INTRODUCTION
The Automatic Biochemistry Analyzer can measure various biochemical indicators by analyzing the blood and other body fluids. Combined with other clinical data, its comprehensive analysis is helpful to diagnose diseases, evaluate organ functions, identify disease factors, and determine future treatment standard.
Leave Your Message:
Write your message here and send it to us.