Nkọwapụta
Nlereanya Obere Oriri akpaaka Hematology Analyzer AMAB24
Ihe nyocha nke hematology na akpaaka AMAB24 atụmatụ:
5-akụkụ dị iche iche, 29 parameters, 1 3D mgbasa gram, 3 2D gbasaa gram na 3 histograms.
Mgbasa Laser akụkụ nke akụkụ atọ jikọtara ya na usoro ịcha mmiri, cytometry eruba elu
Kọmpat, dị ike na ọnụ ọnụ
Naanị 20ul sample olu
Ruo 80 sample kwa elekere
Ụdị ọnụọgụ 3: ọbara venous, ọbara capillary na preluted
Ikike ịkpọpụta ihe atụ na-adịghị mma
Nnukwu ikike nchekwa: ruo 100,000 samples
Auto Hematology Analyzer AMAB24Nkọwa nka nka:
Ụkpụrụ
Usoro mgbochi maka ọnụ ọgụgụ WBC, RBC, PLT
reagent na-enweghị cyanide maka ule haemoglobin
Flow Cytometry (FCM), Semi-conductor Laser gbasasịa, usoro esiji kemịkal, ọwa Basophil nọọrọ onwe ya
Oke
WBC, Lym%, Lym#, Mon%, Mon#, Neu%, Neu#,Eos%, Eos#, Bas%, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW- SD, PLT, PDW, MPV, PCT, na 4 nyocha parameters P-LCC , P-LCR, LIC%, LIC #, ALY%, ALY#
1 3D na 3 2D gbasasịa gram na 3 histogram
Ntinye aka
Ruo 80 sample kwa elekere
Ihe atụ olu
Ọbara dum: 20ul
Nke mbu: 20ul
Ụdị ule
CBC CBC+DIFF
Arụmọrụ
Na-ebughachi Linearity Repeatability
WBC ≤0.5% ≤2.0% (4-15×109/L) 0.00-300×109/L
RBC ≤0.5% ≤1.5% (3.5-6.0×10)12/L) 0.00-8.5 × 1012/L
HGB ≤0.5% ≤1.5% (110-180g/L) 0-250g/L
PLT ≤1.0% ≤4.0% (150-500×10)9/L) 0.00-3000×109/L
Ike Nchekwa data
Ruo 100,000 nsonaazụ gụnyere ọnụọgụ na ozi eserese
Nzikọrịta ozi
LAN ọdụ ụgbọ mmiri na-akwado HL7 protocol
Environment na-arụ ọrụ
Okpomọkụ: 15oC-30oC
iru mmiri: 30-85%
Nrụgide ikuku: 70-106 kPa
Achọrọ ike
AC 100-240V, ≤300VA, 50/60Hz
Akụkụ na ibu
620mm(L) × 620mm(W) × 535mm(H)
Arọ: 56kg