H7c82f9e798154899b6bc46decf88f25eO
H9d9045b0ce4646d188c00edb75c42b9ek

Low Sample Consumption Auto Hematology Analyzer AMAB24

Short Description:

Product Name:Low Sample Consumption Auto Hematology Analyzer AMAB24
Latest Price:

Model No.:AMAB24
Weight:Net weight: Kg
Minimum Order Quantity:1 Set Set/Sets
Supply Ability:300 Sets per Year
Payment Terms:T/T,L/C,D/A,D/P,Western Union,MoneyGram,PayPal


Product Detail

Product Tags

Quick Details

Principles

Impedance method for WBC, RBC, PLT counting

Cyanide-free reagent for Hemoglobin test

Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye method, independent Basophil channel

Packaging & Delivery

Packaging detail:Standard export package
Delivery detail:within 7-10 workdays after receipt of payment

Specifications

Low Sample Consumption Auto Hematology Analyzer AMAB24

Auto Hematology Analyzer AMAB24 Features:

  5-part differentiation, 29 parameters, 1 3D scatter gram, 3 2D scatter grams and 3 histograms

  Tri-angle semi-conductor Laser scatter combined with chemical dye method, advanced flow cytometry

  Compact, powerful and affordable

  Only 20ul sampling volume

  Up to 80 samples per hour

  3 counting modes: venous blood, capillary blood and prediluted

  Capability to flag abnormal samples

  Large storage capacity: up to 100,000 samples

Auto Hematology Analyzer AMAB24 Technical Specifications:

Principles

Impedance method for WBC, RBC, PLT counting

Cyanide-free reagent for Hemoglobin test

Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye method, independent Basophil channel

Parameters

WBC, Lym%, Lym#,Mon%, Mon#,Neu%, Neu#,Eos%, Eos#, Bas%, Bas#,  RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PDW, MPV, PCT,and 4 research parameters  P-LCC , P-LCR, LIC%, LIC#, ALY%, ALY#

1 3D and 3 2D scatter grams and 3 histograms

Throughput

Up to 80 samples per hour

Sample Volume

Whole Blood: 20ul

Prediluted: 20ul

Test Mode

CBC   CBC+DIFF

Performance

Carryover                      Repeatability                      Linearity

WBC   ≤0.5%            ≤2.0% (4-15×109/L)            0.00-300×109/L

RBC   ≤0.5%            ≤1.5% (3.5-6.0×1012/L)        0.00-8.5×1012/L

HGB   ≤0.5%            ≤1.5% (110-180g/L)            0-250g/L

PLT    ≤1.0%            ≤4.0% (150-500×109/L)        0.00-3000×109/L

Data Storage Capacity

Up to 100,000 results including numeric and graphical information

Communication

LAN port supports HL7 protocol

Operating Environment

Temperature: 15oC-30oC

Humidity: 30-85%

Air Pressure: 70-106 kPa

Power Requirement

A.C. 100-240V, ≤300VA, 50/60Hz

Dimension and Weight

620mm(L) × 620mm(W) × 535mm(H)

Weight: 56kg

 

 